Diagnostic performance of commercially available enzyme linked immunosorbent assay kits for detection of PR3-ANCA compared with direct immunofluorescence using cut off levels provided by the manufacturer
Test | Provided cut off | Sensitivity | Specificity | |||
---|---|---|---|---|---|---|
CA, C-ANCA; IFT, indirect immunofluorescence technique; IHELI, in-house enzyme linked immunosorbent assay; neg, negative; pos, positive; PR3-ANCA, antineutrophil cytoplasmic antibodies to proteinase 3. | ||||||
IFT (CA) | 1:16 | 73.3 | 98 | |||
IHELI | <2.5 neg, >5 pos | 66.7 | 100 | |||
A | >2.0 | 60.0 | 100 | |||
B | <10 neg, >15 pos | 13.3 | 100 | |||
E | <7 neg, >10 pos | 46.7 | 100 | |||
F | >15 | 26.7 | 98 | |||
G | >3.5 | 60.0 | 100 | |||
H | <0.9 neg, >1.1 pos | 60.0 | 98 | |||
I | >20 | 60.0 | 100 | |||
J | >2 | 60.0 | 100 | |||
K | <20 neg, <30 pos | 60.0 | 100 | |||
L | <10 neg, >20 pos | 40.0 | 100 | |||
M | >15 | 26.7 | 96 |