Dual-energy CT for the diagnosis of gout: an accuracy and diagnostic yield study
  1. Tim Bongartz1,
  2. Katrina N Glazebrook2,
  3. Steven J Kavros3,
  4. Naveen S Murthy2,
  5. Stephen P Merry4,
  6. Walter B Franz III4,
  7. Clement J Michet1,
  8. Barath M Akkara Veetil1,
  9. John M Davis III1,
  10. Thomas G Mason II1,
  11. Kenneth J Warrington1,
  12. Steven R Ytterberg1,
  13. Eric L Matteson1,
  14. Cynthia S Crowson5,
  15. Shuai Leng2,
  16. Cynthia H McCollough2
  1. 1Division of Rheumatology, Department of Internal Medicine, Mayo Clinic, Rochester, Minnesota, USA
  2. 2Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA
  3. 3Department of Orthopedic Surgery, Mayo Clinic, Rochester, Minnesota, USA
  4. 4Department of Family Medicine, Mayo Clinic, Rochester, Minnesota, USA
  5. 5Division of Biomedical Statistics and Informatics, Department of Health Sciences Research, Mayo Clinic, Rochester, Minnesota, USA
  1. Correspondence to Tim Bongartz, Mayo Clinic College of Medicine, Division of Rheumatology, Department of Internal Medicine, Mayo Clinic, 200 1st Street SW, Rochester, MN 55905, USA; bongartz.tim{at}mayo.edu

Abstract

Objectives To assess the accuracy of dual-energy CT (DECT) for diagnosing gout, and to explore whether it can have any impact on clinical decision making beyond the established diagnostic approach using polarising microscopy of synovial fluid (diagnostic yield).

Methods Diagnostic single-centre study of 40 patients with active gout, and 41 individuals with other types of joint disease. Sensitivity and specificity of DECT for diagnosing gout was calculated against a combined reference standard (polarising and electron microscopy of synovial fluid). To explore the diagnostic yield of DECT scanning, a third cohort was assembled consisting of patients with inflammatory arthritis and risk factors for gout who had negative synovial fluid polarising microscopy results. Among these patients, the proportion of subjects with DECT findings indicating a diagnosis of gout was assessed.

Results The sensitivity and specificity of DECT for diagnosing gout was 0.90 (95% CI 0.76 to 0.97) and 0.83 (95% CI 0.68 to 0.93), respectively. All false negative patients were observed among patients with acute, recent-onset gout. All false positive patients had advanced knee osteoarthritis. DECT in the diagnostic yield cohort revealed evidence of uric acid deposition in 14 out of 30 patients (46.7%).

Conclusions DECT provides good diagnostic accuracy for detection of monosodium urate (MSU) deposits in patients with gout. However, sensitivity is lower in patients with recent-onset disease. DECT has a significant impact on clinical decision making when gout is suspected, but polarising microscopy of synovial fluid fails to demonstrate the presence of MSU crystals.

Keywords
  • Gout
  • Dual energy CT scanning
  • accuracy

This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 3.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/3.0/

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Supplementary materials

  • Supplementary Data

    This web only file has been produced by the BMJ Publishing Group from an electronic file supplied by the author(s) and has not been edited for content.

    Files in this Data Supplement: