Poor accuracy and interobserver reliability of knee arthroscopy measurements are improved by the use of variable angle elongated probes
  1. S P Oakley1,
  2. I Portek1,
  3. Z Szomor2,
  4. A Turnbull2,
  5. G A C Murrell2,
  6. B W Kirkham3,
  7. M N Lassere1
  1. 1Department of Rheumatology, St George Hospital, University of New South Wales, Sydney, Australia
  2. 2Department of Orthopaedics, St George Hospital
  3. 3Department of Rheumatology, Guy's and St Thomas' Hospital Trust, London, UK
  1. Correspondence to:
    Dr S Oakley, Department of Rheumatology, St George Hospital, Gray St, Kogarah, NSW 2217 Australia;
    z8425086{at}student.unsw.edu.au

Abstract

Objectives: (a)To determine the accuracy and reliability of arthroscopic measurements of cartilage lesion diameter in an artificial right knee model; (b) to determine whether the use of a set of variable angle elongated probes improves performance; and (c) to identify other sources of variability.

Methods: Ovoid “lesions” were drawn on the five cartilage surfaces of four plastic knees models. Two observers assessed these 20 lesions arthroscopically, measuring two diameters in orientations parallel and orthogonal to the probe. Observer 1 (orthopaedic surgeon) and observer 2 (arthroscopic rheumatologist) made two sets of measurements, firstly with the conventional probe and five months later with the variable angle elongated (VAE) probes. The knees were disarticulated to determine true lesion diameter.

Results: Observer 1 had negligible bias and good accuracy regardless of orientation or probe type. Observer 2 demonstrated both bias and poor accuracy using the conventional probe. Both improved using VAE probes. Poor interobserver reliability with conventional probes also improved using VAE probes. Major sources of variability could be traced to the probe type, the characteristics of the operator, and the orientation of the lesion in relation to the probe; the lesion location itself did not cause variability.

Conclusions: Variation in accuracy and poor interobserver reliability of measurements with conventional methods of cartilage lesion diameter measurement improved when specially designed measurement probes were used. Arthroscopic measurements performed as well as most clinical and radiographic measures. These findings have important implications for the use of arthroscopy as an outcome in multicentre trials where arthroscopists have different levels of experience.

  • arthroscopy
  • knee
  • validity
  • reliability
  • ANOVA, analysis of variance
  • ICC, intraclass correlation coefficients
  • SDD, smallest detectable difference
  • VAE, variable angle elongated (probes)