Table 3

Intrareader ICC and SDC for MRI, ultrasonography and CT

Intrareader ICC for status scoresSDC (SDC%)
0 Months6 Months12 Months0–6 Months0–12 Months
MRI erosion score0.930.970.983.0 (6.4)2.5 (5.1)
MRI synovitis score0.870.640.492.3 (11.6)2.5 (12.6)
MRI bone oedema score0.440.950.912.4 (7.9)2.6 (7.8)
MRI erosion volume0.980.990.9924 mm3(2.8)32 mm3(3.0)
Ultrasonography erosion score0.920.950.951.8 (5.1)2.1 (6.0)
Ultrasonography grey-scale synovitis score0.910.890.962.4 (6.7)2.6 (7.2)
Ultrasonography power Doppler score0.940.990.971.2 (6.0)1.0 (4.9)
CT erosion score0.970.960.960.3 (0.5)0.7 (1.2)
CT erosion volume0.990.990.9919 (1.6)17 (1.5)
  • ICC, intraclass correlation coefficient; SDC, smallest detectable change; SDC%, SDC as a percentage of actually obtained maximum score.