Article Text

PDF
THU0555 Validation of a Portable Device (RE.MO.TE.) for the Hand Functional Assessment in Patients with Chronic Rheumatic Diseases
  1. M. Piga1,
  2. I. Tradori1,
  3. D. Pani2,
  4. A. Cauli1,
  5. A. Dessì2,
  6. P. Garau1,
  7. G. Porru1,
  8. G. Barabino2,
  9. A. Vacca1,
  10. L. Raffo2,
  11. A. Mathieu1
  1. 1Rheumatology Unit, University of Cagliari AOU, Monserrato
  2. 2Dept. of Electrical and Electronic Engineering, University of Cagliari, Cagliari, Italy

Abstract

Background Patients with inflammatory rheumatic diseases characterized by a chronic progressive course, such as Rheumatoid Arthritis (RA) and Systemic Sclerosis (SSc), develop a disability of the hand which requires an adequate assessment and therapeutic intervention through a pharmacological treatment and a proper rehabilitation program. The Re.Mo.Te. device consists of sensorized aids, allowing the execution of hand rehabilitation exercises and the extraction of relevant parameters (motion, force, pressure, speed of execution) from the analysis of several repetitions of the same movement, performed in real rehabilitation exercises typically prescribed to RA and SSs patients (1).

These data are part of a larger project called Re.Mo.Te. Recovery Movement and Telemonitoring for rheumatology patients with disabilities of the hand.

Objectives To validate portable device Re.Mo.Te for the evaluation of hand disability in patients with RA and SSc.

Methods Ten RA patients (10F, age 56.9 ± 13.1 years) and 10 SSc patients (9F, age 56.3 ± 10.4 years) underwent an assessment of hands function through the device Re.Mo.Te. and through the measurement of: 1) Range Of Movement (ROM) of the wrists and 2) grip strength (pinch and grip) of the hand and fingers, 3) hands and fingers extension. The HAQ and the algo-functional index of Dreiser were administered to both RA and SSc patients while the Hamis test was administered exclusively to SSc patients.

The Spearman’s correlation coefficient was applied to test the validity of the results obtained with the Re.Mo.Te. device and to compare them with traditional assessment; p values ​​<0.05 were considered significant.

Results All patients had a reversible hand disability (HAQ = AR: 1.5 ± 0.82; SSc: 1.4 ± 0.65 - Dreiser’s index = AR: 15.7 ± 6.0; SSc: 15.0 ± 5.7). The Spearman’s correlation coefficients between the traditional hand functional assessment and the Re.Mo.Te. device assessment were high (0.67-0.79) and statistically significant for the measurement of grip strength (p <0.0001) and overall extent (p <0.001). The ROM results showed significant correlation with the global finger extension (p <0.01) and the speed of fingers extension (p <0.001) abilities evaluated by Re.Mo.Te.

Conclusions The Re.Mo.Te. device is a valuable aid for the assessment of hand disability in chronically ill subjects requiring a quantitative evaluation for the proper set up of a personalized rehabilitation program.

References

  1. D. Pani, G. Barabino, A. Dessì, M. Piga, I. Tradori, A. Mathieu, L. Raffo “An integrated Portable Device for the hand functional assessment in the clinical practice” Biomedical Engineering Systems and Technologies 2012 - Revised Selected Papers, J. Gabriel et al. Eds., in press

Disclosure of Interest None Declared

Statistics from Altmetric.com

Request permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.